About cookies on this site Our websites require some cookies to function properly (required). In addition, other cookies may be used with your consent to analyze site usage, improve the user experience and for advertising. For more information, please review your options. By visiting our website, you agree to our processing of information as described in IBM’sprivacy statement. To provide a smooth navigation, your cookie preferences will be shared across the IBM web domains listed here.
Publication
IEEE Transactions on Magnetics
Paper
Measurement of magnetization on an obliquely deposited film by using 3-dimensional vector VSM
Abstract
An obliquely deposited film is experimentally analyzed by using a newly developed three-dimensional vector Vibrating Sample Magnetometer (VSM). Since the M-H loops of the magnetization components orthogonal to the applied field change their signs when the applied field direction crosses the easy or hard direction, detailed information about the magnetic anisotropy can be obtained from the behavior of the magnetization components. Analysis reveals that the system does not have simple uniaxial anisotropy, but has both oblique uniaxial anisotropy and shape anisotropy. © 1991 IEEE