About cookies on this site Our websites require some cookies to function properly (required). In addition, other cookies may be used with your consent to analyze site usage, improve the user experience and for advertising. For more information, please review your options. By visiting our website, you agree to our processing of information as described in IBM’sprivacy statement. To provide a smooth navigation, your cookie preferences will be shared across the IBM web domains listed here.
Publication
TRANSDUCERS 2003
Conference paper
Mass-loaded cantilevers with suppressed higher-order modes for magnetic resonance force microscopy
Abstract
We describe the design, fabrication and testing of mass-loaded cantilevers for electron-spin magnetic resonance force microscopy. These single-crystal silicon cantilevers are designed to have large gaps in their thermal mode spectra so as to reduce thermal noise near the electron-spin Rabi frequency. Each cantilever typically consists of a 2 μm thick mass suspended at the end of a 0.1 μm thick hinge. The fabrication process starts with an SOI wafer followed by selective silicon epitaxy, cantilever patterning, and backside release. The focus of the process is on precise thickness control and material homogeneity. We will present characterization results for these cantilevers at room and low temperatures and discuss the impact of these devices on magnetic resonance force microscopy measurements.